Shop - Diarad 360r

Diarad 360r

General Features

  • 21 parameters
  • 10.4 inch LCD touch screen
  • Up to 60 samples per hour
  • Excellent repeatability and wider linearity range
  • Powerful capability to flag abnormal cells
  • Large storage capacity: up to 50,000 samples
  • Recommended or customizable decision rules for re-exam abnormal samples
  • Both thermal printer and USB external printer are supported
  • No need flushing fluid
Category:
  • Technical Specification
  • Other Advantages
  • Download Manual

Rigorous background requirement:

Parameter Background
WBC ≤0.3×109/L
RBC ≤0.03×1012/L
HGB ≤1 g/L
HCT ≤0.5%
PLT ≤10×109/L

 

Low carryover:

Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.5%
HCT ≤0.5%
PLT ≤1.0%

 

Good precision - stable result:

Parameter (testing range) Whole blood precision (CV) Pre-diluted precision (CV)
WBC (7 - 15.0 ´ 109/L) ≤2.0% ≤4.0%
RBC (3.5 - 6.0 ´ 1012 / L) ≤1.5% ≤3.0%
HGB (110 - 180 g/L) ≤1.5% ≤3.0%
MCV (70 - 120 fL) ≤1.0% ≤2.0%
PLT (150 - 500 ´ 109 / L) ≤4.0% ≤8.0%

 

Wide linearity range:

Parameter Linearity range
WBC (0.00~300.00)×10⁹/L
RBC (0.00~8.50)×1012/L
HGB 0~250g/L
PLT 0~3000×10⁹/L
HCT 0~67%

 

Good Co-relation:

Parameter a b r
WBC 0.97 ≤ a ≤ 1.03 ≤ 0.3 ≥ 0.99
RBC 0.97 ≤ a ≤ 1.03 ≤ 0.1 ≥ 0.99
HGB 0.97 ≤ a ≤ 1.03 ≤ 2 ≥ 0.98
MCV 0.97 ≤ a ≤ 1.03 ≤ 1.5 ≥ 0.98
PLT 0.97 ≤ a ≤ 1.03 ≤ 10 ≥ 0.95

 

★ Low blood consumption: Whole Blood-9 µL, Prediluted-20 µL.

★ User friendly operation: Easily operating software.

★ Long reagent expiration date: 2 years expiration date for all reagents.

★ Easy maintenance: Daily maintenance is performed simply by soaking of probe cleaner during shutdown.